AAOS 2026: long-term performance of BIOLOX®delta & BIOLOX®forte ceramic bearings in THA
Kim et al.: 20- to 30-year bilateral outcomes in younger patients
- 200 patients/400 hips, mean age 46.3 years
- 28mm Duraloc Alumina CoC or Duraloc CoHXLPE
- Each patient received CoC in one hip and CoHXLPE in the other
- Mean follow-up 23.1 years (up to 30 years)
Outcomes
- No osteolysis detected in either group (radiographs and CT)
- Mean liner penetration of 0.041 mm/year
- 96.5% stem survivorship & 97% cup survivorship in both groups
- No fracture of ceramic heads or acetabular liners reported
- Squeaking reported in 32 cases, no cases revised
In this younger, active cohort, both CoC and CoHXLPE showed high long-term survivorship and low wear over 2+ decades of follow-up.
Park et al.: national registry data on BIOLOX®delta ceramic
- 15,299 primary THA patients (2003-2013) received a BIOLOX®delta head at 10-year follow-up
- 14,255 received DoD, 1,044 DoHXLPE
Key findings
- 10-year survivorship: DoD 96.5%, DoHXLPE 95.3%
- Approximately 2× higher risk of PJI in the first year in the DoHXLPE group vs. DoD
- Revision (any cause): 38% higher risk in DoHXLPE group
- No significant differences in dislocation or periprosthetic fracture
In this large, real-world cohort, both DoD & DoHXLPE constructs were associated with high 10-year survivorship, with BIOLOX® DoD bearings showing higher survivorship and lower observed risks of infection-related and all-cause revision.
Seddigh et al.: post-approval study of 36mm BIOLOX®delta CoC THA
- 170 patients/172 hips
- Mean follow-up 5.6 years
- Kaplan-Meier survivorship: 95.3% at 12 years
- Revision rate: 4.7%
- No revisions due to dislocation, infection, or aseptic loosening
Clinical & bearing-related findings
- Harris Hip Score improved from 51.3 pre‑op to 92.5 at 5 years
- No ceramic head fractures during clinical follow-up
- One intraoperative liner fracture reported
- Noise events occurred in ~21.5%, mostly transient – no impact on revision risk or patient satisfaction
- No association between BMI & revision risk
These post-approval data reinforce the midterm safety, durability, and functional performance of large-head DoD bearings in real-world clinical use.
Please check for regulatory approval in your country.